English:Accuracy, precision and reproducibility in measurements - Science Year 10
Introduction
Two teams measure the same reference block. One produces nearly identical readings that lie well above the reference. The other reports more spread but a mean close to the reference. Which should you trust, and what evidence would make you change your mind? Use the original records in this lesson to distinguish accuracy from precision, decide which comparisons test repeatability and reproducibility, and find plausible sources of random variation and systematic bias. Keep the raw data throughout. Your final judgement should say which measurements you compared, which changes were controlled and what remains uncertain.
Compare a cluster with an independent reference
A tight group of readings and a good match to a reference answer different questions. Precision describes agreement among repeated results under the stated conditions. Accuracy concerns closeness to a true quantity value, which is usually unknown. A suitable independently established reference lets you judge agreement within stated limits. In what follows, the quantity to measure is the length between two marked flat ends of a teaching block at room conditions. We take 100.0 mm as the reference and use a reference uncertainty of 0.1 mm for this model. The cases and numbers are invented so you can check each claim.
| Set | Readings in millimetres | Mean in millimetres | Range in millimetres |
|---|---|---|---|
| Instrument A | 102.3, 102.4, 102.5, 102.4 | 102.4 | 0.2 |
| Instrument B | 99.3, 100.6, 100.4, 99.7 | 100.0 | 1.3 |
The range is the maximum minus the minimum. A has a smaller range, so the four results are more precise by this measure. Its mean is 2.4 mm above the reference. B's mean matches the reference, but the individual readings range from 0.7 mm below to 0.6 mm above. Describe what the mean and range separately support. A match to a reference in one small set does not prove instrument B is accurate for other lengths or that its spread comes from one known cause.

Resolution is a property of the measurement system. A digital instrument with displays at steps of 0.1 mm can show a finer change than a ruler marked every 1 mm. That alone does not establish that a digital reading is closer to the reference. An instrument can display a small step and still have a zero offset, an incorrectly aligned contact or a misleading unit setting. State the measurand and the units before making numerical comparisons. If you only know the repeated readings, you can describe their spread, but you need other evidence before you claim the readings are close to a reference.
While watching, note how clustering relates to random error. Using only the two original sets above, explain which statement is about precision and which is about agreement with the block's value. Then identify what the ruler image cannot tell you about either digital instrument.
Test an offset without losing the raw readings
Random variation and systematic bias require different checks. A random component can produce unpredictable differences between repeats, for example if you now and then start a watch slightly early or late. An offset can shift readings in one direction, for example if a sensor reads above zero when its input is zero. In real investigations both effects can occur together. Repeating measurements and reporting a mean can reduce the effect of independent random variation on that mean. It does not remove a constant offset. Repeating a biased measurement may give an impressive-looking cluster around the wrong value.
An independent length check of instrument A gives this model calibration record. The standards were selected before the block under test was measured. A further standard near 120 mm is held back for a check of the correction.
| Reference length in millimetres | Instrument A reading in millimetres | Reading minus reference in millimetres |
|---|---|---|
| 50.0 | 52.4 | +2.4 |
| 100.0 | 102.4 | +2.4 |
| 150.0 | 152.4 | +2.4 |
| 120.0, held-back check | 122.5 | +2.5 |
A plausible model over the tested interval is a +2.4 mm offset. Under this model subtract 2.4 mm from each A reading and keep the uncorrected readings beside the corrected ones. The original block values become 99.9, 100.0, 100.1 and 100.0 mm. Their mean becomes 100.0 mm and the range stays 0.2 mm. For the held-back standard, 122.5 minus 2.4 gives 120.1 mm, which is 0.1 mm above the 120.0 mm reference. It supports the correction in this example but also shows why you should document residual differences instead of promising exact results. There is no evidence here that the correction applies outside the checked range or after the instrument is dropped, reset or used with different contact pressure.

If a class repeats a time measurement, an unpredictable delay can broaden the spread. If the equipment always begins timing after the same unintended trigger delay, every trial can share a bias. Record a clear start and end event, inspect the raw observations and compare the timer with a separately checked reference when feasible. Keep an unusual reading in the original table. Annotating a demonstrable mistake, such as a recorded accidental stop, is stronger than deleting a value solely because it changes the answer. When the reason is uncertain, compare summaries with and without it and explain how that affects the conclusion.
Test agreement when a person or method changes
A repeat test and an independent comparison address different sources of confidence. You test repeatability by using the same method, operator, instrument and relevant conditions and seeing whether runs agree closely. You test reproducibility by deliberately changing relevant conditions, for example by asking a separate team to use its own equipment or an alternative suitable method, while keeping the measurand equivalent. Agreement after changing conditions extends the evidence. Two exact-looking values from one instrument do not tell you whether another instrument or observer would get a similar result.
Three sets below are fictional readings of the same strip. Before comparing, the class sets a rule for this exercise: a difference of no more than 0.5 mm between group means counts as acceptable agreement for the stated task. It is not a universal scientific cut-off. Both instruments display readings to 0.1 mm, but there is no independently established strip length in this comparison.
| Team and procedure | Readings in millimetres | Mean in millimetres |
|---|---|---|
| P: measured the two marked flat ends | 150.1, 150.2, 150.3 | 150.2 |
| Q: initially measured one end including a removable tab | 151.0, 151.1, 151.2 | 151.1 |
| Q: repeated with its own instrument after both ends were defined | 149.9, 150.0, 150.1 | 150.0 |
All three sets have a range of 0.2 mm. Initial team means differ by 0.9 mm. Each has tight repeats, yet the initial comparison fails the agreement rule and the endpoints are not equivalent. In the final comparison, the mean difference is 0.2 mm, which meets the stated rule with the same endpoint definition. Keep the initial record and label the change in procedure. The comparison supports agreement between these two teams using the revised endpoint definition. Without an independent reference it cannot establish the actual strip length or rule out an error common to both groups.
Before a new investigation, specify exactly what you will measure, how you will record one observation, which conditions stay the same and which independent check you will use. Where possible, collect the measurements without revealing the first team's result. That removes one way an expectation can influence the second recording. Label the uncertainties you actually have evidence for, and separate a proposed cause from a demonstrated one. If the trial data disagree, report that outcome and choose a test that can help discriminate among endpoint definition, instrument bias and variable observation conditions. Use the course on presenting observations to decide how best to show both means and individual readings.
Interactive Tasks
Quiz: Test Your Knowledge
Which statement describes set A before correction? (Its readings cluster more tightly but its mean is farther from the reference) (!Its mean is closer to the reference than set B) (!Its smaller display step proves the reference is wrong) (!Its range shows that all random and systematic errors are zero)
What is the range of the four original set B readings? (1.3 mm) (!0.6 mm) (!0.7 mm) (!0.2 mm)
What information does a 0.1 mm display step provide directly? (The size of the displayed increments) (!The confirmed accuracy at all lengths) (!The result of the held-back check) (!The amount of bias caused by an observer)
What happens when a justified constant correction is subtracted from each reading? (The mean shifts while the range stays the same) (!The range becomes zero but the mean stays the same) (!Only the largest reading changes) (!The instrument automatically becomes accurate everywhere)
What is the corrected reading for the held-back 120.0 mm standard? (120.1 mm) (!124.9 mm) (!119.9 mm) (!120.0 mm by definition)
Which comparison most directly tests repeatability? (The same operator repeats a measurement with the same setup) (!Another team measures an unrelated object) (!Two people compare instruments that have never been used) (!A student chooses a reading after seeing the expected result)
What did the initial gap between teams P and Q support? (The initial means fail the agreed difference rule and describe different endpoints) (!Small within-team ranges prove they measured the same thing) (!Team Q was definitely using a faulty instrument) (!The later repetition must be discarded)
Which evidence best tests an instrument for a systematic effect across a range? (Compare several independent standards along that range) (!Repeat one unknown reading until it stops changing) (!Choose the smallest observed reading) (!Check how many decimal places appear on screen)
Which situation can cause random timing variation? (The observer sometimes reacts early and sometimes late) (!The timer adds exactly one second each time) (!A worksheet always converts seconds using the same wrong factor) (!The equipment always starts at a fixed wrong reference)
How should you handle an unusually large reading with an uncertain cause? (Keep it visible and report how it affects the conclusion) (!Erase it and retain only the preferred mean) (!Always replace it with the expected value) (!Retake readings until every value matches)
Memory Game
| Accuracy | Closeness of a measurement to the true quantity value |
| Precision | Agreement within a set of readings taken under stated conditions |
| Resolution | Smallest change that the measurement system can display or distinguish |
| Repeatability | Agreement when the same operator repeats the method under the same conditions |
| Reproducibility | Agreement when relevant observers or equipment change while the quantity remains equivalent |
Drag and Drop
| Held-back standard | Check whether an already chosen correction also works on independent evidence |
| Persistent offset | Compare with a suitable reference before correcting the data |
| Scatter in timings | Improve the timing method and retain the spread across repeats |
| Different end marks | Define the same quantity before comparing two teams |
| Unexplained extreme value | Preserve the raw result and test whether the conclusion is sensitive to it |
Crossword Puzzle
| Accuracy | Closeness to an independently supported value |
| Precision | Quality represented by clustering under stated conditions |
| Resolution | Smallest indicated measurement step |
| Calibration | Comparison that relates an instrument indication to reference values |
| Range | Highest reading minus lowest reading |
| Bias | Systematic tendency for readings to shift in a direction |
LearningApps
Optional extra practice. Choose an appropriate activity with your teacher. If the external collection is unavailable, use the native activities and independent tasks in this course.
Cloze Text
Open Tasks
Easy
- Label qualities: Write a sentence each about the precision and the agreement with reference of sets A and B.
- Read an instrument scale: Point to the unit you would use on a ruler with two scales. Explain why a finer-looking scale does not establish accuracy.
- Compute a spread: Use set B to show the subtraction needed to find its range. Give the answer with units.
- Choose a fair repeat: Specify three things to keep the same when one observer repeats a block measurement.
Standard
- Retain a correction ledger: Make a table with original reading, correction and corrected reading for set A. Summarise what changed in the mean and range.
- Write an independent protocol: Set up a paper comparison for two teams measuring the same strip. State its endpoints, how results are kept independent and a justified agreement rule for the task.
- Test a timing claim: Compare unpredictable delays and an exactly repeated trigger delay. For each, propose one targeted improvement and say how you would judge whether it helped.
- Report a changed method: Write a short correction log for team Q. Explain how the initial and revised values support different conclusions.
Advanced
- Check a non-constant error: A device reads 50.5 mm for a 50.0 mm standard and 151.5 mm for a 150.0 mm standard. Challenge a proposed constant correction and design one check within the interval.
- Review unusual data: A new set contains 99.9, 100.0, 100.1 and 104.0 mm. Report means with and without the last value. Propose a reasoned follow-up without deciding that the last number is erroneous from its size alone.
- Plan a blind cross-check: Design a two-method test in which no team sees the other team's results first. Give an example of a shared bias that could remain even if the results agree.
- Separate evidence and judgement: Write a short report for a class choosing an instrument. Make one decision using your data, mark an unverified assumption and describe a further calibration check.
Learning Check
- Compare two distributions: Use set A and set B to identify the stronger evidence for precision and for agreement with the 100.0 mm reference. Justify the distinction with two numbers.
- Evaluate the held-back check: Apply the calibrated correction to the 122.5 mm reading. Compare it with the independent value. Explain what the result shows about applying the same correction to the block data.
- Diagnose a shared effect: Three observers use the same stopwatch that starts exactly 0.4 s late on every trial. They use a separate reliable finish event. What is the error direction for the elapsed time, and what would reduce it?
- Reconcile the two-team records: Compute both differences between means for teams P and Q. Compare each with the agreed threshold and explain why the revised comparison is more informative.
- Judge an unexpected value: In the four-value set 99.9, 100.0, 100.1 and 104.0 mm, what does including the last value do to the mean? Give a decision rule that keeps the evidence auditable.
- Transfer to a digital balance: A balance shows 0.3 g when empty and gives nearly identical values for a reference mass in repeated trials. What can you infer and what independent check is still needed?
Worked checks and assessment guidance
- A is more precise by range, 0.2 mm against 1.3 mm. B's mean is 100.0 mm while A's mean is 102.4 mm, so B's mean is closer to the reference in this particular comparison. That does not establish the accuracy of each B reading.
- 122.5 minus 2.4 equals 120.1 mm, 0.1 mm above 120.0 mm. The result supports an approximate constant correction in the tested region. Residual error and uncertainty remain and the data do not justify extrapolation.
- Starting late gives an elapsed time 0.4 s shorter if the finish is unaffected. The observations can agree despite sharing that systematic effect. Correct the trigger or independently measure and document the delay. More repeats alone cannot remove it.
- The initial difference is 0.9 mm, outside the 0.5 mm rule, and endpoint definitions differ. The revised difference is 0.2 mm and both teams use the same endpoints, so the revised comparison supports agreement for that task. It does not establish the true length.
- With all four, the mean is 101.0 mm. Without the last point it is 100.0 mm. Keep both calculations and record the circumstances of the last trial. Exclude it from a chosen summary only with a defensible reason and show that choice and its effect.
- The indication suggests a zero offset and the repeats show precision under the stated conditions. Check the balance against one or more independent reference masses at relevant loads and report whether the offset remains constant. The repeat readings alone cannot establish accuracy.
Learning evidence
Keep the two original measurement sets, your signed correction ledger, a held-back check and a report comparing both teams before and after the endpoints were defined. Include evidence of spread and agreement separately. State one bounded conclusion supported by the data and one possible explanation that still needs a different observation.
OER and Wikipedia
- Department for Education, National curriculum in England: Science, Key Stage 4 Working Scientifically.
- AQA GCSE Combined Science, practical assessment objective WS 3.7.
- Institute of Physics, Physical quantities guidance notes.
- Scale ruler, 1 foot and 300 mm.jpg, Kaviyadharshini Selvakumar, cropped and rotated on Commons by SCHolar44, CC BY-SA 4.0. Used as supplied by Commons.
- Stoppuhr.jpg, Ulfbastel, public domain. Used as supplied by Commons.
- GCSE Science Revision: Precision, Freesciencelessons.
- GCSE Working Scientifically: Repeatability and Reproducibility, Freesciencelessons.
- All datasets, comparisons and assessment prompts in the course are original examples.
Linked learning areas
aiMOOC Projects
Schulfach+


aiMOOCs


aiMOOC Projekte


MOOCwiki · Deutsch
Nach dem Lernen ist vor dem Lernen
Entdecke direkt den nächsten Lernkurs. Weitere Inhalte erscheinen, wenn Du weiter nach unten scrollst.
Zur MOOCwiki-HauptseiteMediathek
Mediathek wird aus dem Wiki geladen ...
Keine passenden Inhalte gefunden. Bitte ändere Suche oder Filter.
NEWSLernweltNOAH fragen